Description
Fischer DE15 – DP with integrated Operating pressure Transmitter 
dual pressure monitoring for non-aggressive gases including cryogenic applications
Make: Fischer Messtechnik (Germany)
Model: DE15
Fischer DE15 is uniquely designed, state-of-art differential pressure measuring transmitter with integrated operating pressure measuring system. The integrated pressure sensor records operating pressure of the system independent of differential pressure measuring system. Thereby, DE15 offer two independent measurement and output signal – one for differential pressure and other for operating pressure.
It can be used for non-corrosive gases and vapours including cryogenic applications.

Working Principle:
Differential pressure measuring system
DE15 measuring system consists of two pressure chambers separated by a diaphragm. Differential pressure applied on either one of the chambers leads to proportional displacement of the diaphragms. This movement is transferred to an inductive displacement transducer and converted to electrical analog signal by downstream electronics.
Operating pressure measuring system
A ceramic sensor with DMS bridge is installed on (-) side of the device. The current pressure changes the form of the ceramic membrane and leads to a proportional change in resistance of DMS bridge. This change of resistance is converted to 4-20mA analog output signal.

Key Features:
- Configurable measuring range
- Configurable parameters via LCD Display, keypad
- Integrated operating pressure measurement up to 40 bar
- Suitable for non-corrosive gaseous media and vapour
- Corrosion resistant field housing
- Oil and grease free for oxygen applications
Key Specifications:
- Measuring range: 0 ~ 160 mbar to 0 ~ 2.5 bar
- Maximum static pressure: 40 bar
- Design: Differential pressure measurement, with or without integrated operating pressure measurement (25 bar or 40 bar)
- Electrical output: differential pressure & operating pressure — 4-20mA 2-wire connection
- Maximum pressure load: one-sided overpressure-proof up to the rated pressure of the measuring system, resistance to under-pressure on the (+) and (-) side
- Linearity: < 1% Full scale
- Ambient temperature: -20 °C to +70 °C
- Admissible media temperature: -20 °C to +70 °C
- Characteristic curve: linear
- Pressure connection: inner thread G 1/4 cutting ring screw connection for 8 or 10 mm pipe
- Measuring membrane: 1.4305 NiBe for differential; Al2O3 for operating pressure
- Protection: IP65 as per EN 60529, with integrated ceramic sensor
Application areas:
- Filter condition monitoring
- Flow measurement and control
- Pump Control
Application areas:
- Nuclear power plants
- Manufacturing and chemical plants
- Marine, shipping application
- Onshore and offshore area
- Petrochemical, solvent industry
- Sewage, effluent treatment plants